Katarina Marković, Radmila Milačič, Janja Vidmar, Stefan Marković, Katja Uršič, Martina Nikšić Žakeljc, Maja Cemazar, Gregor Sersa, Mojca Unk, Janez Ščančar
Journal of Trace Elements in Medicine and Biology, Volume 57, January 2020, Pages 28-39.
Abstract
Monolithic chromatography using convective interaction media (CIM) disks or columns can be used in the separation step of speciation analysis. When different monolithic disks are placed in one housing, forming conjoint liquid chromatography (CLC) monolithic column, two-dimensional separation is achieved in a single chromatographic run. Here, we assembled low-pressure (maximum 50 bar) CLC monolithic column, which consists of two 0.34 mL shallow CIM monolithic disks and high-pressure CLC column (maximum 150 bar) from 0.1 mL analytical high performance short bed CIMac monolithic disks. The data from analyses showed that both tested CLC monolithic columns gave statistically comparable results, with the low-pressure CLC column exhibiting better resolving power and robustness. Low-pressure CLC column exhibited greater potential than high-pressure CLC column, and can be thus recommended for its intended use in speciation analysis of metal-based biomolecules.