Urh Černigoj, Jana Vidič, Ana Ferjančič, Urša Sinur, Klemen Božič, Nina Mencin, Anže Martinčič Celjar, Pete Gagnon, Aleš Štrancar
Electrophoresis, September 2021
Abstract:
Elution of strong and weak anion exchangers with sodium chloride gradients is commonly employed for analysis of sample mixtures containing different isomers of plasmid DNA. Gradient elution of a weak anion exchanger (diethylaminoethyl, DEAE) in the presence of guanidine hydrochloride (Gdn) roughly doubles resolution between open-circular (oc) and supercoiled (sc) isomers. It also improves resolution among sc, linear, and multimeric/aggregated forms. Sharper elution peaks with less tailing increase sensitivity about 30%. However, elution with an exclusively-Gdn gradient to 900 mM causes more than 10% loss of plasmid. Elution with a sodium chloride gradient while maintaining Gdn at a level concentration of 300 mM achieves close to 100% recovery of sc plasmid while maintaining the separation improvements achieved by exclusively-Gdn elution. Corresponding improvements in separation performance are not observed on a strong (quaternary amine) anion exchanger. Other chaotropic salts do not produce a favorable result on either exchanger, nor does the inclusion of surfactants or EDTA. Selectivity of the DEAE-Gdn method is orthogonal to electrophoresis, but with better quantification than agarose electrophoresis, better quantitative accuracy than capillary electrophoresis, and resolution approaching capillary electrophoresis.