On May 12th, the biaseparations.com website will be retired and migrated tosartorius.com.Learn moreabout our combined offering today!

Miniaturized Monolithic Disks for Immunoadsorption of Cardiac Biomarkers from Serum

N. Delmotte, U. Kobold, T. Meier, A. Gallusser, A. Strancar, C. G. Huber

Anal Bioanal Chem (2007) 389:1065–1074

Immunoadsorbers based on 2.0 × 6.0 mm i.d., epoxy-bearing, methacrylate-based monolithic disks were developed in order to target myoglobin and N-terminal pro-natriuretic peptide (NT-proBNP), two biomarkers involved in cardiovascular disease. In both cases, antibodies were successfully coupled to the polymeric disk material. The developed immunoadsorbers permitted the selective isolation of myoglobin and NT-proBNP from human serum. Myoglobin was successfully isolated and detected from serum samples at concentrations down to 250 fmol μL-1. However, the affinity of the antibodies was not sufficient for the analysis of low-concentration clinical samples. Frontal analysis of anti-NT-proBNP disks revealed the ability of the immunoadsorber to bind up to 250 pmol NT-proBNP, which is more than sufficient for the analysis of clinical samples. Anti-NT-proBNP disks showed good stability over more than 18 months and excellent batch-to-batch reproducibility. Moreover, anti-NT-proBNP disks permitted the isolation of NT-proBNP at concentrations down to 750 amol μL−1 in serum, corresponding to concentrations of strongly diseased patients. Using reversed-phase trapping columns, the detection of NT-proBNP eluted from immunoadsorbers by mass spectrometry was achieved for concentrations down to 7.8 fmol μL-1.

Purchase full article