On May 12th, the biaseparations.com website will be retired and migrated tosartorius.com.Learn moreabout our combined offering today!
2005

Fast diagnosis of different infections is a crucial for a successful medical treatment. For diagnosis of certain diseases, separation of IgG and IgM in human serum is required to prevent interference or competing. This is usually achieved by adding adsorbent containing antihuman antibodies to the sample. Incubation from half to one hour is needed to achieve the complete removal of the antibody.

A quicker way to achieve the removal of antibody would be the use of a chromatographic support with specific ligand, which selectively binds the antibody. For example, a Protein G column can be used for removal of IgG. This is faster, but also much more expensivfe way of removing IgG's.

CIM Convective Interaction Media stationary phases represent a novel generation of stationary phases for liquid chromatography. Because of their monolithic structure, being designed for the separation and purification of macromolecules, they exhibit a higher dynamic capacity for very alrge molecules in comparison to traditional stationary phases, combined with much shorter process time that further result in a decreased loss of the biologic activity.

In this work, we present low price ligands (coupled to CIM chromatographic support), which can be used for efficient separation of IgG and IgM antibodies.

Attachments

Full view

2004

By using a combination of two CIM® tube monolithic columns, OH and DEAE chemistry, we were able to successfully purify plasmid DNA from bacterial culture without using RNase. Purified plasmid DNA is very pure, since common contaminants, such as proteins, genomic DNA, endotoxins and RNA were under the detection limit. The scale up units produced according to cGMP standard are already used for the purification of plasmid DNA for gene therapy purposes on industrial scale.

Attachments

Full view

Immobilized Metal-Affinity Chromatography (IMAC) is a separation technique primarily intended for the purification of proteins with exposed histidine tags. Technique uses covalently bound chelating compounds on chromatographic supports to entrap metal ions, which serve as affinity ligands for various proteins. Iminodiacetic acid (IDA), nitrilotriacetic acid (NTA), carboxymethylated aspartic acid (CM-Asp), and N,N,N’-tris(carboximethyl) ethylenediamine (TED) are chelating compounds, most often used to entrap metal ions, such as Cu2+, Ni2+, Zn2+, Co2+ etc.

Convective Interaction Media CIM® is a monolithic support, which provides high rates of mass transfer at low pressure drops. It has been shown that CIM® supports are very efficient for the separation of large molecules, such as proteins and DNA (1). Recent publication has proved that CIM IMAC column can be used for separation of histidine containing peptides (2). Since efficient separation of large molecules is one of the main advantages of CIM® support, purification of His-tagged recombinant proteins on CIM IMAC column should be not only feasible but also simple, fast and efficient.

Attachments

Full view

Membrane bound heterotrimeric guanine-nucleotide proteins (G-proteins) are the important components of the cellular signal transduction cascade. They are GTPases which cycle between an inactive and an active configuration by catalysing the exchange of GTP for GDP bound to G subunit. In our study we investigated separation of high affinity GTP'S binding proteins (G-proteins) from plasma membrane of porcine brain by HPLC using CIM® (Convective Interaction Media) supports. CIM® supports proved to be an efficient tool for cytosolic protein separation on second or minute time scale. No study of separation of membrane bound proteins by CIM® supports have been done so far.

Attachments

Full view

Tissue plasminogen activator (t-PA) is serine protease which converts plasminogen into plas-min dissolving the major component of blood clots, fibrin. So, it can be extremely useful in clinical practice to help curing of heart attack victims. The most available way protein producing is genetic engineering where separation and purification of goal protein are one of the important steps in protein producing process.

Recently developed High performance monolithic disk chromatography, HPMDC, seems to be a very attractive way for study quantitative affinity parameters of recombinant proteins with different ligands as well as for protein separations and purifications. High process speed prevents the denatura-tion due to temperature and solvents influence. The better mass transfer mechanism (convection rather than diffusion) allows to consider only the biospecific reaction as time limiting.

It is known that plasminogen, which is the natural substratum for t-PA, can be successfully used as affinity ligand to separate t-PA from cellular media. However, the use of synthetic ligands for affinity chromatography is more preferable due to their higher stability and lower total cost.

Attachments

Full view

2003

The only four drugs approved for the clinical treatment of Alzheirner’s Disease (tacrine. rivastigmine, donepezil and galantamine) are acetylcholinesterase inhibitors which act by maintaining high levels of acetylcholine at the muscarinic and nicotinic receptors in the central nervous system. Human acetylcholinesterase (HuAChE) represents a widely studied target enzyme and it is still object of research for the development of new drugs as enzyme inhibitors.

In a previous paper we reported the immobilisation of AChE on a silica based chromatographic column (50 x 4.6 mm 1.0.) The yeld of immobilization and the stability of the AChE-IMER were considered satisfactory, but some problems arose. The length of the IMER and the large amount of enzyme covalently bound to the chromatographic support resulted in catalysis product long elution times and some inhibitors aspecific matrix absorption with delayed enzyme activity recovery. In order to avoid these complications and considering the high rate of AChE enzymatic reaction. we decided to reduce the dimension of the solid support for immobilization, hence the amount of immobilized enzyme, by selecting a monolithic matrix disk (12 x 3 mm I.D.).

CIM® (Convective Interaction Media) monolithic supports (Biaseparations. Lubiana) represent a novel generation of stationary phases used for liquid chromatography, bioconversions, and solid phase synthesis. As opposed to individual particles packed into chromatographic columns, CIM® supports are cast as continuous homogeneous phases and provide high rates of mass transfer at lower back pressure.

In the present work a CIMK disk with immobilised human recombinant acetylcholinesterase (HuAChE-ClM® Disk) was developed. The activity of immohilised enzyme, the long term stability and reproducibility were tested. HuAChE-CIM® disk was applied as an immobilised enzyme micro-reactor (micro-IMER) in on-line HPLC system for inhibitory potency determination of known AChE inhibitors.

Full view

Gene therapy which is becoming more and more important in human health care requires the purification of high molecular mass compounds, so called nanoparticles (e. g. viruses and plasmids). The method of choice to ensure proper purity would be chromatography.

Most of the chromatographic supports available on the market at the moment can not follow the requests for such work due to low binding capacity for large molecules, limitation with regards to the time of the separation process and requests for CIP (cleaning in place) and SIP (sanitation in place).

Monolithic supports represent a new generation of chromatographic supports. In contrast to conventional particle supports, where the void volume between individual porous particles is unavoidable, these supports consist of a single monolith highly interconnected with larger and smaller open flow-through channels. Due to the structure, molecules to be separated are transported to the active sites on the stationary phase by convection, resulting in very short separation times. This is especially true for large molecules.

In this work we will present the use of monolithic supports for the separation of different nanoparticles on analytical and preparative scales. It will be shown that monolithic supports can overcome the limitations of particle-based supports for the analytics and isolation of big molecules and represent a major step towards the safe and efficient purification or production of nanoparticles.

Attachments

Full view

Plasmids are episomes that have been recognized in few eukaryotic and most prokaryotic species. Some plasmids are excellent genetic vectors and they have been widely used in gene manipulation and recombinant DNA technology for a long time. In recent years plasmids were intensively used for gene therapy purposes (1). Most often purification starts with the cells harvest followed by alkaline lysis step in which ribonuclease A (RNase) is typically used. After that, plasmid DNA can be precipitated and used directly or can be further purified by different methods (2). Currently, several chromatographic methods, such as ion-exchange, size exclusion, affinity, and hydrophobic chromatography, have been demonstrated in plasmid purification (3). Until now a limited number of small scale purification methods without use of RNase were published. Convective Interaction Media CIM® is a monolithic chromatographic support for which has been shown that is very efficient for the separation of large molecules, such as proteins, DNA and viruses (4).

Attachments

Full view

Traces of DNA in RNA samples represent impurities that could affect results of mRNA quantification and cDNA synthesis. In most cases, the DNA impurities in RNA samples are removed using enzyme deoxyribonuclease (DNase), which specifically breaks down DNA. In order to avoid the addition of DNase into the analyzing sample, the use of immobilized DNase on solid support is recommended. Because of the DNA size, very few supports available on the market enable efficient interaction between immobilized enzyme and DNA.

In recent years a new group of supports named monoliths was introduced. Because of enhanced exchange between mobile and stationary phase separation and bioconversion processes are significantly accelerated. Therefore also the efficiency of DNA removal using immobilised enzyme might be competitive to the degradation with free enzyme.

Attachments

Full view

Plasmids are episomes that have been recognized in few eukaryotic and most prokaryotic species. Some plasmids are excellent genetic vectors and they have been widely used in gene manipulation and recombinant DNA technology for a long time. In recent years plasmids were intensively used for gene therapy purposes (1).Most often purification starts with the cells harvest followed by alkaline lysis step in which ribonucleaseA (RNase) is typically used. After that plasmid DNA can be precipitated and used directly or can be further purified by different methods (2).Currently, several chromatographic methods, such as ion-exchange, size exclusion, affinity, and hydrophobic chromatography, have been demonstrated in plasmid purification (3). Until now a limited number of small scale purification methods without use of RNase were published. Convective Interaction Media CIM®is a monolithic chromatographic support for which has been shown that is very efficient for the separation of large molecules, such as proteins, DNA and viruses (4).

Attachments

Full view

The only four drugs approved for the clinical treatment of Alzheimer’s Disease (tacrine, rivastigmine, donepezil and galantamine) are acetylcholinesterase inhibitors which act by maintaining high levels of acetylcholine at the muscarinic and nicotinic receptors in the central nervous system. Human acetyicholinesterase (HuAChE) represents a widely studied target enzyme and it is still object of research for the development of new drugs as enzyme inhibitors.

In a previous paper il] we reported the immobilisation of AChE on a silica based chromatographic column (50 x 4.6 mm I.D.) The yield of immobilization and the stability of the AChE—IMEN were considered satisfactory, hut some problems arose. The length of the IMER and the large amount of enzyme covalently bound to the chromatographic support resulted in catalysis product long elution times and some inhibitors aspecific matrix absorption with delayed enzyme activity recovery. In order to avoid these complications and considering the high rate of AChE enzymatic reaction, we decided to reduce the dimension of the solid support for immobilization, hence the amount of immobilized enzyme, by selecting a monolithic matrix disk (12 x 3 min I.D.).

CIMa (Convective Interaction Media) monolithic supports (Bia Separations, Ljubljana) represent a novel generation of stationary phases used for liquid chromatography, bioconversions, and solid phase synthesis. As opposed to individual particles packed into chromatographic columns, CIM supports are cast as continuous homogeneous phases and provide high rates of mass transfer at lower back pressure.

In the present work a CIM® disk with immobilised human recombinant acetylcholinesterase (HuAChECIM€ Disk) was developed. The activity of immobilised enzyme, the long term stability and reproducibility were tested. HuAChECIM disk was applied as an immobilised enzyme micro-reactor (micro-IMER) in on-line HPLC system for inhibitory potency determination of known AChE inhibitors.

Attachments

Full view

2002

The progress in gene-therapy and DNA vaccination leads to a growing demand of therapeutic applicable plasmid DNA (pDNA). To guarantee the supply for the clinical trials and finally for the market new pDNA production processes, which meet all regulatory requirements, have to be developed. Conventional small scale techniques can not easily be transferred to the manufacturing scale (technical reasons and safety considerations). We developed a generic large scale process for highly purified plasmids “free” of bacterial contaminants which works without enzymes, detergents (except SDS during the cell lysis) and organic solvents.

Attachments

Full view

Most commonly plasmids are manufactured by fermentation of E. coli. In the cells several isoforms of the plasmid are generated: supercoiled (sc), open circular (oc) and linear as well as dimeric forms. After alkaline lysis plasmids are accompanied in solution by genomic DNA (gDNA), RNA, proteins and other cell compounds [1]. In addition to these impurities, the plasmid isoforms have to be separated efficiently in order to get a final product containing > 95 % of ccc form [2]. Chromatographic resins used in biotechnology are usually designed for the separation of polypeptides, providing only low capacity for polynucleotides (< 1 mg/mL).

In this work we present an optimised purification step for large scale purification of therapeutic applicable pDNA, based on an alternative chromatography resin (CIM Convective Interaction Media®).

Attachments

Full view

2001

We have developed a screening procedure for peptide ligands for affinity chromatography on the same monolithic support. CIM® monolithic columns used conventionally for analytical and preparative separation of proteins and polynucleotides were minimized to fit into 96 well solid phase extraction plates. Peptide synthesis and screening were performed on the same format using a vacuum manifold for liquid throughput.

Attachments

Full view

The development of new chromatographic supports with the aim to improve their chromatographic, hydrodynamic and mechanical properties is continually going on.

CIM Convective Interaction Media® monolithic columns represent a new chapter in every mode of the chromatography. Monolithic columns consist of a single piece of a highly porous polymer with a bimodal pore size distribution, forming flow-through channels [1]. Since all of the mobile phase flows through the pores, molecules to be separated are transported to the active sites by convection [2]. Therefore, the entire analysis can be completed in a very short time.

In this work, the performance of novel semi-preparative CIM® RP-SDVB disk monolithic column for separating proteins and peptides has been investigated. Since the column length in the case of gradient separations commonly used for large molecules, does not play a significant role, CIM® RP-SDVB disk monolithic column are extremely short, typically of only 3 mm. The effect of decreasing column length on the resolution under the conditions of a linear gradient has been presented.

Finally, a 1 minute purification of oligodeoxynucleotide from the synthetic mixture has been performed.

Attachments

Full view

Gene therapy which is becoming more and more important in human health care requires the purification of high molecular mass compounds, so called nanoparticles (e. g. viruses and plasmids). The method of choice to ensure proper purity would be chromatography.

Most of the chromatographic supports available on the market at the moment can not follow the requests for such work due to low binding capacity for large molecules, limitation with regards to the time of the separation process and requests for CIP (cleaning in place) and SIP (sanitation in place).

Monolithic supports represent a new generation of chromatographic supports. In contrast to conventional particle supports, where the void volume between individual porous particles is unavoidable, these supports consist of a single monolith highly interconnected with larger and smaller open flow-through channels. Due to the structure, molecules to be separated are transported to the active sites on the stationary phase by convection, resulting in very short separation times. This is especially true for large molecules.

In this work we will present the use of monolithic supports for the separation of different nanoparticles on analytical and preparative scales. It will be shown that monolithic supports can overcome the limitations of particle-based supports for the analytics and isolation of big molecules and represent a major step towards the safe and efficient purification or production of nanoparticles.

Attachments

Full view

CIM® (Convective Interaction Media) supports represent the fourth generation of chromatographic supports monoliths. These glycidyl methacrylate based monolithic columns are commercially available under the name CIM®. In contrast to conventional porous particles the morphology of the CIM® supports is characterised by a single monolithic unit that contains pores, opened on both sides. These pores are highly interconnected forming a flow-through network. Thus the whole mobile phase is forced to run through these open pores, therefore, the mass transfer between stationary and mobile phase is based on convection rather than on diffusion. This transport mechanism enables very fast separations and purifications of components as well as flow-unaffected resolution and dynamic binding capacity. The latter is especially important for the purification of molecules on preparative level where the productivity is essential. One of the key features of monolithic units is their pore size distribution that should enable low backpressure at high throughputs together with high specific surface area, needed for high binding capacity. In this work the effect of different parameters on binding capacity of CIM® monolithic columns is presented.

Attachments

Full view

There is a rapidly increasing development of new materials in the field of chromatographic supports stimulated by the need to achieve fast and reliable separation of different solutes.

The generally used chromatographic supports are based on beads packed in the columns. Although these stationary phases have been continuously improved over the last decades, there are still some limitations present. The absence of flow within the pore matrix of particles, slow diffusional mass transfer of solutes, in some cases high back pressure and laborious handling represent the major hindrances.

Development of monolithic materials is a chronicle of efforts to overcome problems of packed particles. Monolithic separation media, made in one piece, contain only flow-through pores, which significantly augment the mass transfer based on convection. This enables use of high mobile phase velocities along with low back pressures and therefore fast separations without decrease of resolution.

This report presents the preparation of glycidylmethacrylatestearylmethacryate- ethylenedimethacrylate and styrene-divinylbenzene monoliths. The porous structure of the support obtained will be discussed. Finally, a few examples of separation of proteins and small molecules with monolithic CIM C-18 and CIM SDVB disks in reversed-phase chromatography will be displayed.

Attachments

Full view

Ion exchange chromatography is one of the most commonly used techniques for the purification and separation of polar samples such as minerals in water or charged biomolecules (Figure 1).

The technique is based upon reversible binding of the charged species to an oppositely charged group that is attached to an insoluble matrix. A quantitative measure of an ion exchanger’s ability to take up exchangeable counter-ions is its capacity, which strongly influence support properties and can be measured by potentiometric titration with a strong acid or base [2]. However, the time to achieve the ion exchange equilibrium (the stationary state the potentiometric titration is based upon) is very long [3]. Consequently, a new method to measure the total ionic capacity of anion exchange resins is being developed.

Attachments

Full view

CIM Convective Interaction Media® are polymer-based monolithic supports which were introduced for chromatographic analyses, in-process control, solid phase extraction, and purification of target biomolecules, both on an analytical and on a preparative scale 1, 2. CIM supports perform high-resolution separations within seconds. This is predominantly due to the convective mass transport of the biomolecules between the mobile and stationary phases and the very low dead volume of the separation unit. One of the main concerns in the last few years was the batch-to-batch reproducibility of the monoliths during manufacturing and the possibility of using the monolithic supports for validated analytical methods. The batch-to-batch reproducibility in product preparation as well as its stability during analytical work should fulfill all the requirements for a validated analytical method. To demonstrate that this is possible, we have selected one complex example – the determination of impurities in immunoglobulins (IgGs) where a multidimensional, so called CLC (Conjoint Liquid Chromatography), approach combining the ion exchange and affinity chromatography was needed to properly analyze the sample.

Therefore, two CIM Protein G disks and one CIM QA disk were placed in series in one housing. Binding conditions were optimized in a way that the IgGs were bound to the CIM Protein G disks while Transferrin and Albumin were separated on a CIM QA disk. A complete separation of all three proteins was achieved in five minutes.

Attachments

Full view