CIM® EDA-8f mL Tube Monolithic Column for Ligand Immobillization
 |
Cat. No.: 430.5116
|
General Description:
CIM
® EDA immobilization columns are amine activated monolithic supports obtained by reacting the native epoxy groups with ethylene diamine. CIM
® EDA monoliths can be used for affinity chromatography or bioconversion by immobilizing proteins, peptides and other ligands through a crosslinking reaction with a suitable bifunctional reagent, e. g. glutaric dialdehyde.
Learn more on CIM® Tube technology >>
Features and Benefits:
- Flow unaffected resolution
|
Extremely fast separations speed up method development and decrease costs. Speed also enables in-process control and analytics.
|
- High dynamic binding capacity
|
High binding capacity, even for large molecules, boost productivity even to 10-fold! |
- High throughput at low pressure drop
|
High flow rates can be achieved at moderate pressure drops- no need for expensive equipment (pumps).
|
- Same material and design accross all products.
|
Same material and design accross all sizes, enable easy scale-up from analytical, semi-prep and industrial scale. |
Dimensions and Operating Conditions:
Monolith Dimensions
- Outer diameter, Do: 15.0 mm
- Inner diameter, Di: 6.5 mm
- Length, L: 56.0 mm
- Bed volume, V: 8.0 mL
Operating Conditions
- Flow rates: up to 400 mL/min (uav = 1340 cm/h)
WARNING: Reversing the flow direction can seriously damage the column!
- Pressure: up to 20 bar (2 Mpa)
- Temperature: 4 °C (39 °F) to 50 °C (122 °F)
- Recommended Ph:
working range 2–13*
cleaning-in-place 1–14* (e. g. 1 M NaOH)
* Valid for the CIM
® EDA Disk.
Note: After immobilization, the working and CIP range depend on the stability of the coupled ligand!